The absolute measurement of ¹³¹I radioactivity for γ-ray nuclide standardization

AUTHORS
Korawee Saejia

Nuclear Scientist, Office of Atoms for Peace (OAP), Thailand

Tanapol Dachviriyakij

Radiation Physicist, Office of Atoms for Peace (OAP) and Korea Research Institute of Standards and Science (KRISS), Thailand

Noppalerk Thongdeelert

Nuclear Scientist, Office of Atoms for Peace (OAP), Thailand

Kanokwan Luansri

Radiation Physicist, Office of Atoms for Peace (OAP), Thailand

Sumalee Ninlapruk

Head of Ionizing Radiation Metrology Section, Office of Atoms for Peace (OAP), Thailand

Vithit Pungkun

Expert on Radioactivity Measurement, Office of Atoms for Peace (OAP), Thailand

5
Views
0
Downloads
MEDIA
Download PDF
DETAILS
Theme 3. Health and Life Sciences
Conference 150th anniversary of the Metre Convention — From Units to the Universe
Year 2025
Location Palais des congrès de Versailles, Versailles, France
Publisher Bureau International des Poids et Mesures (BIPM)
Language EN
ABSTRACT
The National Standard Radioactivity Laboratory (NSRL) of the Office of Atoms for Peace (OAP), the regulatory authority for radiation and nuclear energy safety in Thailand, aims to improve and maintain primary activity standardization results. To this end, the NSRL determined calibration factors for a 4π (re-entrant) ionization chamber for ¹³¹I for participation in an Interlaboratory Proficiency Testing (IPT) Program in 2025. This program allows the 21 participating laboratories across the country performing activity measurements to compare their results using radionuclide calibrators. Reference values were measured using the 4π ionization chamber coupled to a Keithley 6517B electrometer controlled by a homemade LabVIEW program. The chamber was calibrated using a ¹³¹I solution source standardized by a 4πβ Pressurized Proportional Counter (PPC)-γ coincidence system, with ⁶⁰Co as a tracer. While the ionization chamber's response curve has been traceable to the National Metrology Institute of Japan (NMIJ) since 2004, the OAP determined a new calibration factor for certifying the activity of ¹³¹I aqueous sources to ensure accuracy and precision. The ¹³¹I absolute activity measurement yielded 902.89 kBq/g with a range uncertainty of 0.61 kBq/g and a coverage factor of k=2. The new ¹³¹I calibration factor is 6.85 pA/MBq. This value represents a 19.75% difference from the previous calibration factor of 5.72 pA/MBq. For further measurement improvement, key issues related to the chamber system's calibration for gamma emitters and short half-life radionuclides, such as timing, current measurement, background, decay, and ¹⁶⁶ᵐHo check source measurements, are being addressed.
Keywords: 4πβ(PPC)-γ coincidence, Absolute measurement, ¹³¹I, Radionuclide calibrator, Standardization, Traceability
CITE THIS POSTER
Include:
Saejia, K., Dachviriyakij, T., Thongdeelert, N., Luansri, K., Ninlapruk, S., & Pungkun, V.. (2025). The absolute measurement of ¹³¹I radioactivity for γ-ray nuclide standardization. 150th anniversary of the Metre Convention — From Units to the Universe. Palais des congrès de Versailles, Versailles, France. Bureau International des Poids et Mesures (BIPM). https://doi.org/10.59161/IRER6118