Conference150th anniversary of the Metre Convention — From Units to the Universe
Year2025
LocationPalais des congrès de Versailles, Versailles, France
PublisherBureau International des Poids et Mesures (BIPM)
LanguageEN
ABSTRACT
Quantitative MRI (QMRI) provides a measure of tissue physical parameters, such as the transverse relaxation time, T2, which characterises decay of excited transverse magnetisation and is used clinically as a biomarker for diagnosing and monitoring of cardiomyopathy. QMRI therefore has the potential for use in clinical diagnosis and prognosis, but quality assurance is needed to ensure the extraction of reliable and accurate quantitative information. In this study, we examine how changes in factors for the MRI scanner set up and model curve fitting parameters impact on output measurements of T2.
The model that is chosen to fit the MR data may impact on the output measurement of T2. Additionally, the specific radio-frequency pulse sequence chosen, as well as the various parameters which may be set during the scan set-up, may all have an impact on the output measurement. Longer MRI scan times may allow more precise measurements of T2 but are more difficult to transfer into the clinic. Similarly with model fitting, one is limited by the compute time of fitting individual voxels, and the complexity of the model. With more complex models, there is also the risk of over-fitting and introducing bias to the output estimated T2. In practice, there is a compromise between accuracy (bias) and precision (variance) driven by signal-to-noise considerations. Knowing what compromises can be made without impacting strongly on the output T2 estimate is useful for moving from research into clinical MRI.
We use an MRI simulator and digital phantom to systematically investigate the effect of changing individual parameters on the output measured T2 value. We run two full-factorial analysis of variance (ANOVA) sensitivity analyses and show that small changes to factors such as the selected curve fitting method and size of the fitted region of interest (ROI) may have a large impact on the output estimated T2.
The factors investigated for the curve fitting sensitivity analysis include the model, fitting method (voxel-wise or fitting a whole ROI at once), the radius and centring of the ROI. For the sensitivity analysis for the simulated MRI scan set up, the factors investigated were the repetition time, frequency and phase encoding steps, slice thickness and flip angle. We find that the factors with the greatest impact on estimated T2 are the fitting method and size of the chosen ROI, and the frequency and phase encodings in the MR sequence.
Future work with phantom acquisitions and analysis on clinical systems should take care when adjusting these factors as small changes may have a large impact on the output measurement. Standardised acquisition and analysis protocols are highly recommended.
CITE THIS POSTER
Include:
Cooke, E., Ferrero, R., Smith, N., Tatman, B., Clarke, J., Bosnjakovic, A., Cashmore, M., Clarkson, C., Curtis, S., Goldring, J., McCann, A., McGrath, C., Thornton, J., Tofts, P., Manzin, A., McDowell, A., & Hall, M.. (2025). A sensitivity analysis of MRI scanner and model parameters for measuring relaxation times. 150th anniversary of the Metre Convention — From Units to the Universe. Palais des congrès de Versailles, Versailles, France. Bureau International des Poids et Mesures (BIPM). https://doi.org/10.59161/LWFU5754
Cooke, Elizabeth, et al.. “A sensitivity analysis of MRI scanner and model parameters for measuring relaxation times.”. 150th anniversary of the Metre Convention — From Units to the Universe, 2025, Palais des congrès de Versailles, Versailles, France, Bureau International des Poids et Mesures (BIPM). DOI: 10.59161/LWFU5754
Cooke, Elizabeth, et al.. 2025. “A sensitivity analysis of MRI scanner and model parameters for measuring relaxation times.”. 150th anniversary of the Metre Convention — From Units to the Universe, Palais des congrès de Versailles, Versailles, France. Bureau International des Poids et Mesures (BIPM). https://doi.org/10.59161/LWFU5754
@inproceedings{cooke2025a_sensitivity_analys,
author = {Cooke, Elizabeth and Ferrero, Riccardo and Smith, Nadia and Tatman, Ben and Clarke, Jack and Bosnjakovic, Alen and Cashmore, Matt and Clarkson, Cailean and Curtis, Sian and Goldring, Jessica and McCann, Aaron and McGrath, Cormac and Thornton, John and Tofts, Paul and Manzin, Alessandra and McDowell, Amy and Hall, Matt},
title = {A sensitivity analysis of MRI scanner and model parameters for measuring relaxation times},
booktitle = {150th anniversary of the Metre Convention — From Units to the Universe},
year = {2025},
address = {Palais des congrès de Versailles, Versailles, France},
publisher = {Bureau International des Poids et Mesures (BIPM)},
doi = {10.59161/LWFU5754},
url = {https://thebipm150.org/poster-page/10eugqq},
}
TY - CONF
AU - Cooke, Elizabeth
AU - Ferrero, Riccardo
AU - Smith, Nadia
AU - Tatman, Ben
AU - Clarke, Jack
AU - Bosnjakovic, Alen
AU - Cashmore, Matt
AU - Clarkson, Cailean
AU - Curtis, Sian
AU - Goldring, Jessica
AU - McCann, Aaron
AU - McGrath, Cormac
AU - Thornton, John
AU - Tofts, Paul
AU - Manzin, Alessandra
AU - McDowell, Amy
AU - Hall, Matt
TI - A sensitivity analysis of MRI scanner and model parameters for measuring relaxation times
T2 - 150th anniversary of the Metre Convention — From Units to the Universe
PB - Bureau International des Poids et Mesures (BIPM)
PY - 2025
CY - Palais des congrès de Versailles, Versailles, France
DO - 10.59161/LWFU5754
UR - https://thebipm150.org/poster-page/10eugqq
ER -