Conference150th anniversary of the Metre Convention — From Units to the Universe
Year2025
LocationPalais des congrès de Versailles, Versailles, France
PublisherBureau International des Poids et Mesures (BIPM)
LanguageEN
ABSTRACT
We describe advances in the metrology of ultra-small DC electric currents, by which it is now possible to provide traceable calibrations of ammeters at 1 fA with an uncertainty of roughly 1%. Measuring such fantastically small currents (a femtoampere corresponds to roughly 10,000 electrons per second) might seem an esoteric curiosity, but in fact there is an important practical application to air quality monitoring. The concentration of particulates (tiny particles of soot and other material) in air, is measured using special aerosol electrometers which ionise the particles and then measure the current due to the flow of charged particles. The small concentration of particles in ambient air translates into currents in the range from a few femtoamperes to a few hundred femtoamperes, so the calibration of aerosol electrometers requires traceable current sources covering this range. Particulate concentration is a key component of overall air quality, and as this becomes the target of increasingly stringent legislation, the traceability of femtoamp-level currents has assumed increasing importance at NPL and other National Measurement Institutes (NMIs).
The technical advances that made these measurements possible were originally driven by basic research into a new type of current standard. Starting in the 1990s, researchers at NMIs began to experiment with single-electron pumps (SEPs), which generate a current by the controlled movement of single electrons. Such a pump, if free from errors, would generate a current based only on the charge on the electron and the pumping frequency. This research gathered pace in new century, motivated by the upcoming re-definition of the SI base unit ampere and the need for a direct realisation of this unit. However, the pumps generated small currents, and researchers into SEPs grappled with the challenge of comparing and scaling currents well below 1 nanoampere. The response to this challenge was the development of new types of instrumentation, such as the ultrastable low-noise current amplifier (ULCA), originally developed at PTB specifically for single-electron research and now commercially available. Although developed in a research context, the ULCA, and related developments in the metrology of high value resistors, turned out to be powerful tools to deliver lower uncertainties for calibrations at the lab bench. At NPL, the ULCA is used to provide traceable calibration of small DC currents not just for aerosol particulate metrology, but also for radionuclide measurements, and photometry / radiometry. Meanwhile, research into SEPs is ongoing, and in the future, they may be used to directly calibrate low current ammeters.
Thus, fundamental research into improving the realisation of an SI base unit has had an unexpected impact, improving the measurement infrastructure that underpins the quality of life for millions of citizens. This story illustrates the importance of research into basic metrology even if the impact pathway is not obvious at the start.
CITE THIS POSTER
Include:
Giblin, S., & Eastoe, J.. (2025). SI traceability of ultrasmall currents for air quality measurement. 150th anniversary of the Metre Convention — From Units to the Universe. Palais des congrès de Versailles, Versailles, France. Bureau International des Poids et Mesures (BIPM). https://doi.org/10.59161/QAAU4333
Giblin, Stephen, and Jonathan Eastoe. “SI traceability of ultrasmall currents for air quality measurement.”. 150th anniversary of the Metre Convention — From Units to the Universe, 2025, Palais des congrès de Versailles, Versailles, France, Bureau International des Poids et Mesures (BIPM). DOI: 10.59161/QAAU4333
Giblin, Stephen, Jonathan Eastoe. 2025. “SI traceability of ultrasmall currents for air quality measurement.”. 150th anniversary of the Metre Convention — From Units to the Universe, Palais des congrès de Versailles, Versailles, France. Bureau International des Poids et Mesures (BIPM). https://doi.org/10.59161/QAAU4333
@inproceedings{giblin2025si_traceability_of_u,
author = {Giblin, Stephen and Eastoe, Jonathan},
title = {SI traceability of ultrasmall currents for air quality measurement},
booktitle = {150th anniversary of the Metre Convention — From Units to the Universe},
year = {2025},
address = {Palais des congrès de Versailles, Versailles, France},
publisher = {Bureau International des Poids et Mesures (BIPM)},
doi = {10.59161/QAAU4333},
url = {https://thebipm150.org/poster-page/10eugrc},
}
TY - CONF
AU - Giblin, Stephen
AU - Eastoe, Jonathan
TI - SI traceability of ultrasmall currents for air quality measurement
T2 - 150th anniversary of the Metre Convention — From Units to the Universe
PB - Bureau International des Poids et Mesures (BIPM)
PY - 2025
CY - Palais des congrès de Versailles, Versailles, France
DO - 10.59161/QAAU4333
UR - https://thebipm150.org/poster-page/10eugrc
ER -