Scanning Microwave Microscopy for Biomedical Applications

AUTHORS
Bruno Eckmann

Research associate, Federal Institute of Metrology METAS, Switzerland

Johannes Hoffmann

Senior scientist, Federal Institute of Metrology METAS, Switzerland

Sophie de Préville

Research associate, Federal Institute of Metrology METAS, Switzerland

Hung-Ju Lin

Research associate, Eidg. Institut für Metrologie METAS, Switzerland

Markus Zeier

Head of Laboratory, Federal Institute of Metrology METAS, Switzerland

4
Views
0
Downloads
MEDIA
Download PDF
DETAILS
Theme 3. Health and Life Sciences
Conference 150th anniversary of the Metre Convention — From Units to the Universe
Year 2025
Location Palais des congrès de Versailles, Versailles, France
Publisher Bureau International des Poids et Mesures (BIPM)
Language EN
ABSTRACT
In clinical surgery, the rapid diagnosis of biological tissue is crucial. Ideally, surgeons should be able to distinguish between healthy and tumorous tissue in real-time. However, many areas lack measurement capabilities or standardized measurement methods, and current techniques often involve sending samples to a laboratory for quick processing and analysis.

This study explores the potential of Scanning Microwave Microscopy (SMM) for analysing biological tissue. To address the challenges of a realistic measurement environment, such as an operating room, we focus on developing a robust microscope setup and applying the technique to biological samples.

In SMM, a sample's electrical properties are probed with high spatial resolution using near-field microwave signals. These signals are generated by a vector network analyser (VNA) and coupled into a small metallic probe. The signal is reflected depending on the impedance at the probe-sample interface and measured by the VNA. Scanning the tip over the sample produces a 2D map of the sample's electrical properties, such as its complex permittivity, with sub-µm spatial resolution.

This work aims to enhance the robustness of SMM by incorporating through-glass vias (TGVs) as the scanning tip. This innovative approach leverages the mechanical stability and precision of TGVs. The distance between the probe and sample is controlled optically, allowing for rapid and accurate approach control, significantly improving the reliability of SMM measurements. The technique further allows for parallelization which greatly improves measurement speed.

A notable achievement of this study is the successful measurement of a fixated pig brain slice, where the technique provided clear contrast between different brain regions, including gray and white matter. This demonstrates the potential of the enhanced SMM technique for detailed and precise imaging of complex biological tissues.

The outcome of this work is expected to foster innovation in the domain of diagnostic analytics, as it demonstrates the techniques abilities and potential. Further investigations, such as expanding the application scope to skin cancer diagnosis or implementing the technique in a real clinical environment, are planned.

In summary, the work presented here combines techniques from two traditional but complementary metrological fields, nanotechnology and electricity, to advance metrology principles in an area with less established metrological foundations. Furthermore, scanning probe microscopy measurements are delicate and traditionally performed by experts under highly controlled lab conditions. The aim here is to make significant impact in the important field of healthcare by transforming a highly specialized lab procedures into a robust and easy-to-use solution for diagnostic purposes.
CITE THIS POSTER
Include:
Eckmann, B., Hoffmann, J., de Préville, S., Lin, H., & Zeier, M.. (2025). Scanning Microwave Microscopy for Biomedical Applications. 150th anniversary of the Metre Convention — From Units to the Universe. Palais des congrès de Versailles, Versailles, France. Bureau International des Poids et Mesures (BIPM). https://doi.org/10.59161/HYYW1939