Toward standardisation in clinical diagnostics: Investigations into liquid handling robotics to underpin SI traceable, multiple analyte quantification

AUTHORS
Tabatha Hambidge

Lead Researcher, LGC & University of Surrey, United Kingdom

Steve Corless

Science Leader, LGC, United Kingdom

Chloe Deaves

Researcher, LGC, United Kingdom

Giles Drinkwater

Lead Researcher, LGC, United Kingdom

Patrick Sears

Senior Lecturer, University of Surrey, United Kingdom

Heidi Goenaga-Infante

Director of Chemical Metrology and Chief Scientist, LGC, United Kingdom

15
Views
2
Downloads
MEDIA
Download PDF
DETAILS
Theme 3. Health and Life Sciences
Conference 150th anniversary of the Metre Convention — From Units to the Universe
Year 2025
Location Palais des congrès de Versailles, Versailles, France
Publisher Bureau International des Poids et Mesures (BIPM)
Language EN
ABSTRACT
Certified Reference Materials (CRMs) are needed to support standardisation of clinical diagnostics as required by ISO 15189 and the in-vitro diagnostic regulation (IVDR) EU/2017/746. Typically, CRM production takes several months for one or a few analytes. To increase both the number of analytes and speed of CRM production, novel workflows are required that have the capability of traceably measuring multiple analytes, with high accuracy and low uncertainty. Therefore, the automation of calibrator preparation to underpin the characterisation of CRMs by use of liquid handling robots with integrated balances has been investigated.
Double exact matching isotope dilution mass spectrometry (DEM-IDMS) is used as the higher order reference methodology to certify organic compounds in CRMs. A higher order reference method for manually certifying amino acids has been previously developed at the NML, with a standard uncertainty below 1%. This method was translated onto two liquid handling robots with integrated balances, the Agilent 7696A Sample Prep Workbench and the GERSTEL Multipurpose Sampler. Using these robots, automated gravimetric calibrator preparation for the characterisation of a multiple analyte CRM was investigated. The accuracy and precision of the robotic technique was evaluated.
The initial feasibility of both automated systems (robots) to make gravimetric standards was found to be unexpectedly variable. Liquid handling settings such as draw speed, dispense speed, washes, dispense washes, dispense pumps, syringe airgap, and overfill were optimised, however inconsistencies and bias remained. A major difference between analyst preparation and robot preparation was identified as the use of a pipette, as opposed to a syringe piercing a closed vessel with a septum. It was concluded that the variability was likely due to solvent droplets depositing on the septum, during removal of the syringe, which was adding to the weight but not incorporated into the solution, affecting the mass fraction. The use of an automated decapping device was assessed to prevent these piercing errors, and further challenges with evaporation from decapping vessels was explored.
Once the methods on both robotic systems were improved, trial certification of a known CRM (NIST amino acid CRM 2389a in 0.1HCl) and a CCQM certified material (K159 amino acids in plasma) was undertaken, providing comparable mass fractions to those obtained by manual high order IDMS value assignment. This work continues to assess other analytes such as estradiol in plasma and therapeutic drug monitoring in blood with the Euramet MetPart project “Manufacturing of commutable calibrators and quality control materials for standardisation and post-market surveillance of IVD tests” (23IND02 COMET). This presentation will provide an overview of the advances in automation for multiple analyte CRM characterisation as well as highlight key remaining challenges.
CITE THIS POSTER
Include:
Hambidge, T., Corless, S., Deaves, C., Drinkwater, G., Sears, P., & Goenaga-Infante, H.. (2025). Toward standardisation in clinical diagnostics: Investigations into liquid handling robotics to underpin SI traceable, multiple analyte quantification. 150th anniversary of the Metre Convention — From Units to the Universe. Palais des congrès de Versailles, Versailles, France. Bureau International des Poids et Mesures (BIPM). https://doi.org/10.59161/IBBO2534