Advancing dosimetry for radiotherapy in medium-energy x-ray beams: the Italian experience at ENEA-INMRI with a graphite calorimeter

AUTHORS
Susy Toma

PhD student, ENEA_INMRI and University of Cassino and Southern Lazio, italy

Domenico Capriglione

Professor- Dept. of Electrical and Information Engineering, University of Cassino and Southern Lazio, italy

Massimo Pinto

Director-National Institute of Ionizing Radiation Metrology, ENEA_INMRI, Italy

Fausto Arpino

Professor-Dept. of Civil and Mechanical Engineering, University of Cassino and Southern Lazio, Italy

Gianluca Cappadozzi

Technical role- PhD Co-tutor- National Institute of Ionizing Radiation Metrology, ENEA-INMRI, italy

Carlo Di Ianni

Technician, ENEA-INMRI, italy

7
Views
0
Downloads
MEDIA
Download PDF
DETAILS
Theme 3. Health and Life Sciences
Conference 150th anniversary of the Metre Convention — From Units to the Universe
Year 2025
Location Palais des congrès de Versailles, Versailles, France
Publisher Bureau International des Poids et Mesures (BIPM)
Language EN
ABSTRACT
For high-energy photon, electron, proton, and ion beams, radiotherapy dosimetry can be established using absorbed dose to water (D_wDw) standards. However, in the domain of medium-energy x-rays dosimetry, there are only few primary standards of D_wDw available in the world at this time, while calorimetry is comparatively more established at several other radiation modalities employing higher photon energies.
Four primary standards of absorbed dose to water in filtered medium-energy x-ray beams are available at National Metrology Institutes (NMI): three are based on water calorimeters (PTB, Germany; VSL, the Netherlands; LNE-LNHB, France) and one on a graphite calorimeter in a water phantom (ENEA-INMRI, Italy). The IAEA Code of Practice TRS-398 focuses, among other beam qualities, also on medium energy x-rays recommending either an approach based on a direct determination of D_wDw -when this is achievable- or via a conversion to D_wDw starting from a measurement of the quantity air kerma whose primary realization is more established at NMIs.

The aim of this work, carried out at ENEA- INMRI, is to provide data of the calibration factors expressed in terms of D_w (NDwQ) directy determined in water, at 2 g cm−2 and 98 cm Source to Surface Distance (SSD), for a commercial, Farmer-type IBA FC65-G chamber and a PTW 30013 chamber, measured through an optimized metrological performance of the Italian Rx calorimeter. The beam qualities are produced at generating potentials between 100 and 250 kV.

At the same kV x-ray beam qualities, the calibration factors of the same ionization chambers were previously determined in the 16NRM03 RTNORM metrology research normative project [unpublished report], expressed in terms of air-kerma with traceability to the Italian national primary standard at ENEA-INMRI, a free-air- chamber. The comparison of the past results with the re-derived set of calibration coefficients expressed in terms of absorbed dose to water will be shown.
Aim of this study is to improve the accuracy of the determination of the quantity absorbed dose to water in medium energy x-ray beams though a refinement of the calorimeter that constitutes the Italian national standard of absorbed dose to water in medium energy x-rays.

Keywords: graphite calorimetry, calorimetry, medium-energy x-rays, reference dosimetry, primary standard, x-ray dosimetry, measurement uncertainty.
CITE THIS POSTER
Include:
Toma, S., Capriglione, D., Pinto, M., Arpino, F., Cappadozzi, G., & Di Ianni, C.. (2025). Advancing dosimetry for radiotherapy in medium-energy x-ray beams: the Italian experience at ENEA-INMRI with a graphite calorimeter. 150th anniversary of the Metre Convention — From Units to the Universe. Palais des congrès de Versailles, Versailles, France. Bureau International des Poids et Mesures (BIPM). https://doi.org/10.59161/QRZB1173