Optical-based primary standard for the safety and efficacy of medical ultrasound

AUTHORS
Christopher Fury

Senior Scientist, National Physical Laboratory, UK

Srinath Rajagopal

Principle Scientist, National Physical Laboratory, UK

7
Views
0
Downloads
MEDIA
Download PDF
DETAILS
Theme 3. Health and Life Sciences
Conference 150th anniversary of the Metre Convention — From Units to the Universe
Year 2025
Location Palais des congrès de Versailles, Versailles, France
Publisher Bureau International des Poids et Mesures (BIPM)
Language EN
ABSTRACT
Medical ultrasound imaging equipment requires absolute measurement of the generated acoustic fields to demonstrate device safety, but also allowing high output levels to enable efficacy of therapeutic equipment. The 1980s saw the realisation of the first calibrated hydrophones (sensitivity in volts per pascal), which have continued to technologically evolve to this day due to their irreplaceable usefulness in measurement capability both on, and near, the medical production line. The large range of devices on offer have posed a wide range of measurement challenges, be it extended frequency ranges, radio-frequency noise susceptibility, high or low absolute sensitivities, micrometre-level construction defects, or fragility, to name a few. Producing traceable hydrophone calibrations are not trivial, and similarly primary calibration methods are limited to a handful of NMIs. In recent years, the demand for expanded frequency ranges and lower measurement uncertainties from hydrophones has pushed the boundaries of the available absolute calibration methodologies.
NPL has been a leader in metrology for the medical ultrasound sector for several decades, providing secondary customer calibrations over a wide frequency range (0.1 to 60 MHz). For primary calibration, a laser vibrometer is used to measure the motion of an insonified thin film (5 μm thickness), where the measured velocity is proportional to the instantaneous pressure of the acoustic field within the optical focus of the laser beam. This measurement of the true field can then be used to calibrate hydrophones through a substitution method. In 2019 a new optical vibrometer at NPL began a programme to extend the current frequency range and ensure the longevity of traceable measurements. Recent work has led to an improvement of the existing interferometer technique that consequently extended the primary standard’s frequency to cover a wider range of frequencies relevant to medical applications. This covers the main frequencies of interest for broadband hydrophone calibration for acoustic pressure. The method is robust to different sources of acoustic field generation including single frequency quasilinear acoustic fields, finite amplitude distorted fields, and highly broadband narrow-pulse excitation.
We have also realised a new ultrasound generation method in the Photoacoustic source (PAS), that utilises short-duration, rapid energy deposition into an optical absorber, to produce short, single, pulse and highly broadband (100 kHz -100 MHz) ultrasound fields. The method demands lower corrections than existing implementations, particularly regarding spatial averaging, which is crucial for accurately calibrating hydrophones within focused acoustic fields (whose correction can be up to 10 times larger).
Culmination of the new interferometer’s commissioning, along with PAS development, has allowed for recent broadband primary calibration of hydrophones over the course of a single day, where historically such measurements would have required weeks of data collection. This new method will allow for traceability of hydrophone measurements to the primary realisation of the acoustic pascal with improved measurement uncertainty. In future work, the frequency range will be extended further to cover more applications, particularly for high frequency diagnostic scanners, and hitherto difficult key parameters, such as temperature dependence, will be available to study accurately for the first time.
CITE THIS POSTER
Include:
Fury, C., & Rajagopal, S.. (2025). Optical-based primary standard for the safety and efficacy of medical ultrasound. 150th anniversary of the Metre Convention — From Units to the Universe. Palais des congrès de Versailles, Versailles, France. Bureau International des Poids et Mesures (BIPM). https://doi.org/10.59161/PKVD5873